MicroFID
II
Flame Ionization Detector
IPN 074-579-P1A
OPERATING MANUAL
Seitenansicht 0
1 2 3 4 5 6 ... 88 89

Inhaltsverzeichnis

Seite 1 - MicroFID

MicroFID™ IIFlame Ionization DetectorIPN 074-579-P1AOPERATING MANUAL

Seite 2

074-579-P1A 10 WARNING 1. All calibration, maintenance and servicing of this device, including battery charging, must be performed in a safe area a

Seite 3 - Notices

074-579-P1A 11 DECLARATION OF CONFORMITY This is to certify that this equipment, designed and manufactured by: INFICON Inc. Two Technology Place Ea

Seite 4 - Table of Contents

074-579-P1A 12 RoHS: INFICON hereby declares that the above mentioned products are assigned to the classification of Category 9, Industrial T

Seite 5

074-579-P1A 13 1. INTRODUCTION About this Manual This manual provides detailed instructions for the setup, operation and maintenance of the MicroFID

Seite 6

074-579-P1A 14 UNPACKING THE MicroFID II The following accessories are included with your MicroFID II: MicroFID II Instrument CD Manual 70 Liter

Seite 7 - For Your Safety

074-579-P1A 15 Flame Ionization Detector Operation The MicroFID II uses a flame ionization detector for the measurement of combustible organic compo

Seite 8 - Notices and Warnings

074-579-P1A 16 Figure 2. Block Diagram Detector Response The MicroFID II is strictly an organic compound detector. It does not respond to inorganic

Seite 9

074-579-P1A 17 Always secure cylinders before removing the cylinder valve protection cap. Do not drag or roll cylinders. Use a cylinder hand truck

Seite 10 - WARNING

074-579-P1A 18 You must obtain a tank of hydrogen from which you can fill the internal cylinder. When ordering hydrogen, specify ultra-high purity h

Seite 11

074-579-P1A 19 ACCESSORIES Computer The MicroFID II will send information stored in its logged memory or will continuously stream data to a computer

Seite 12

074-579-P1A 2 Warning: Limitation of Liability The ultimate responsibility of the consequences of use of toxic compounds rests with the user. INFIC

Seite 13 - 1. INTRODUCTION

074-579-P1A 20 To connect a charcoal filter: 1. Load the Teflon ferrules into the nut. The nut and ferrules are supplied with the filter. See Figu

Seite 14

074-579-P1A 21 2. USING THE MicroFID II MicroFID II Feature Layout The photograph below shows the basic layout of the MicroFID II Version components

Seite 15

074-579-P1A 22 Using the MicroFID II Battery Charger READ THESE INSTRUCTIONS BEFORE USING THE CHARGER The MicroFID II battery charger is

Seite 16 - Detector

074-579-P1A 23 HOW TO USE THE CHARGER The Charger is started by connecting the battery pack to the charger. Once connected, the indicator LED will

Seite 17

074-579-P1A 24 To Charge the Battery 1. To charge the MicroFID II battery, unscrew the cover for the charging connector cap. 2. Connect the char

Seite 18

074-579-P1A 25 NOTE: The MicroFID II will not operate without the charging connector cap in place. NOTE: The MicroFID II will not operate with the

Seite 19 - ACCESSORIES

074-579-P1A 26 WARNING Hydrogen storage on the metal oxide is an Exothermic Reaction. The cylinder may become hot during the initial fill. Care mu

Seite 20

074-579-P1A 27 Table 2. Charging Time and Usage Time WARNING Use only the MicroFID II Hydrogen Fill Station p/n A1201222 to fill the hydrogen cy

Seite 21 - 2. USING THE MicroFID II

074-579-P1A 28 CAUTION: When filling the MicroFID II Hydrogen Cylinder, ensure that the hydrogen cylinder on/off valve is completely open in the cou

Seite 22

074-579-P1A 29 NOTE: Read through this section before filling the hydrogen fuel cylinder. Store the hydrogen supply tank in a well ventilated area,

Seite 23

074-579-P1A 3 Release History Part Number Release Publication Date MN201103 A 6 September 2011 074-579-P1 A 16 October 2012 Customer Service IN

Seite 24

074-579-P1A 30 Filling the Hydrogen Fuel Cylinder 1. Rotate the output valve to point toward the main cylinder of hydrogen (Filling Position)

Seite 25

074-579-P1A 31 Figure 10. Hydrogen Fuel Cylinder Connected to Filling Station Hydrogen Tank Valve Warning: Do not adjust this regul

Seite 26

074-579-P1A 32 Installing the Hydrogen Fuel Cylinder in the MicroFID II_____________________________ 1. If the hydrogen fuel cylinder retention pla

Seite 27

074-579-P1A 33 WARNING The hydrogen fuel cylinder retention plate MUST be installed on the MicroFID II to retain the intrinsic safety rating. 6.

Seite 28 - Hydrogen Cylinder

074-579-P1A 34 8. The MicroFID II will automatically start with “MicroFID II Setup” on the main display. Simply press the middle red soft key (“meas

Seite 29

074-579-P1A 35 Preparing for Field Operation Field Check List The following items should be carried into the field to reduce or eliminate instrument

Seite 30 - Venting Position

074-579-P1A 36 Operational Check List Before beginning field work, set up and calibrate the MicroFID II for your particular application. Ensure the

Seite 31

074-579-P1A 37 3. USER FUNCTIONS Display The MicroFID II has a graphic display for reporting detected concentrations and to guide you through confi

Seite 32 - 074-579-P1A 32

074-579-P1A 38 main menu Menu Function Displays Figure 12. The MicroFID II Main M

Seite 33

074-579-P1A 39 MicroFID II Setup Menu Audio ON Audio OFF Pump ON Pump OFF Method 21 Logging Off Location Interval Optional Air ON Optional Air OFF

Seite 34

074-579-P1A 4 Table of Contents FOR YOUR SAFETY ...

Seite 35

074-579-P1A 40 Figure 15. Shut Down MicroFID II Menu Figure 16. View Menu Shut Down MicroFID II Shut down in process Press ‘next’ soft

Seite 36

074-579-P1A 41 main menu Keys Fixed ON/OFF Key The ON/OFF key is used to both power on

Seite 37 - 3. USER FUNCTIONS

074-579-P1A 42 enter password Numeric Value, Duration, Time and Date Entry In cases wher

Seite 38

074-579-P1A 43 MicroFID II Display Description Flame Icon with line through it Flame has been extinguished and needs to be relit TVOC high level al

Seite 39 - MicroFID II Setup Menu

074-579-P1A 44 enter password User Interface – Basic Menu The MicroFID II is designed fo

Seite 40

074-579-P1A 45 1/02/09 location 14:45 1/02/09

Seite 41 - 074-579-P1A 41

074-579-P1A 46 1/02/09 method 21 14:45 1/02/09 metho

Seite 42 - Figure 7: Edit Mode Display

074-579-P1A 47 1/02/09 method 21 14:45 1/02/09 method 21

Seite 43 - Table 5. System Alerts

074-579-P1A 48 meth 21 status user to compare the Leak Definition with the Corrected value. Once the c

Seite 44 - 074-579-P1A 44

074-579-P1A 49 enter password The MicroFID II has the ability to change log modes withou

Seite 45 - 45.6 ppm

074-579-P1A 5 3. USER FUNCTIONS ...

Seite 46 - 1.72 ppm

074-579-P1A 50 To delete data logged in the MicroFID II 1. Press the MENU key. 2. Use the DOWN ARROW key to choose “Data Log Options”, and then pr

Seite 47 - 100 ppm

074-579-P1A 51 d) Response Factor = e) Alarm Level = f) Formula Weight = g) CR LF = Carriage Return and Line Feed are accomplished by pressing th

Seite 48 - 074-579-P1A 48

074-579-P1A 52 MicroFID II Setup Functions MicroFID II Setup functions are used to select the MicroFID II features (See Figure 14). There are fou

Seite 49 - Enter new INTERVAL (secs)

074-579-P1A 53 To turn on the pump: 1. Press the MENU soft key, “MicroFID II Setup”, press SELECT, “Pump”, press SELECT, “On”, then press SELECT. T

Seite 50 - 074-579-P1A 50

074-579-P1A 54 enter password Clock The Clock function is used to set both the current d

Seite 51 - 074-579-P1A 51

074-579-P1A 55 Date Format The date format can be expressed either as MM/DD/YYYY, or as DD/MM/YYYY. Using either the Up or Down arrow soft key, ch

Seite 52 - 074-579-P1A 52

074-579-P1A 56 memory slot 1 Compound Selection and Response Factors The response of a fla

Seite 53 - 074-579-P1A 53

074-579-P1A 57 Edit H2 Timer The H2 Timer is used to count down the number of hours of use remaing in the hydrogen fule cylinder. Each time the hydr

Seite 54 - Enter current time

074-579-P1A 58 MicroFID II Calibration Calibration Technical Description The MicroFID II will always detect all ionizable compounds present in a sa

Seite 55 - 074-579-P1A 55

074-579-P1A 59 Methane in air is recommended as span gas. However, Hexane or Propane can also be used as span gas. The choice is made from the Cali

Seite 56 - 074-579-P1A 56

074-579-P1A 6 5. ROUTINE MAINTENANCE ... 74 Ma

Seite 57 - 074-579-P1A 57

074-579-P1A 60 6. Open the regulator by turning the valve counter clockwise, until the ball is approximately 1/8” (3 mm) from its rest position.

Seite 58 - 074-579-P1A 58

074-579-P1A 61 Alternative Calibration with Gas Bag Preparing the Calibration Gas Bag and the Zero Air Bag Connecting the MicroFID II with calibrat

Seite 59

074-579-P1A 62 4. Loosen the knurled nut on the reducing union of the bag adapter. NOTE: Do not remove the nut from the union, as the Teflon ferrul

Seite 60 - 074-579-P1A 60

074-579-P1A 63 select span gas 9. Disconnect the gas bag from the gas bag adapter and empty the

Seite 61 - 074-579-P1A 61

074-579-P1A 64 connect zero zero cal zero too high 3. Connect the suppl

Seite 62 - 074-579-P1A 62

074-579-P1A 65 edit span connect span spam level 6. The MicroFID II the

Seite 63 - Propane Span Gas

074-579-P1A 66 calibrate span span error span completed 8. The unit w

Seite 64 - 074-579-P1A 64

074-579-P1A 67 upon completion of calibration, the unit will revert to the User Mode edit Interval timing screen again. If the user wishes to go int

Seite 65 - 074-579-P1A 65

074-579-P1A 68 4. The following screen will appear: Connect the MicroFID II to your zero air supply and open the valve to the zero air supply. Th

Seite 66 - 074-579-P1A 66

074-579-P1A 69 8. This screen will automatically appear: Press next once you attach the first span gas. 9. The MicroFID II will automatically en

Seite 67 - 074-579-P1A 67

074-579-P1A 7 For Your Safety Strictly follow the Instructions for Use Any use of this instrument requires a full understanding and strict adheren

Seite 68 - 074-579-P1A 68

074-579-P1A 70 12. Attach span gas as instructed. 13. The MicroFID II will automatically enter span gas 2. 14. Once the last calibration st

Seite 69 - 074-579-P1A 69

074-579-P1A 71 TARE selection It is not unusual that, under normal operating conditions, the zero setting on the MicroFID II may drift below zero. T

Seite 70 - 074-579-P1A 70

074-579-P1A 72 Enabling Blue Tooth Communication Users of Computer/PDA datalogging software programs will need to perform the following steps in or

Seite 71 - 4. WIRELESS COMMUNICATION

074-579-P1A 73 Select Real Time Data output ON, then select the Back option 2 times to enter the top level menu or select Measure to resume direct m

Seite 72 - 074-579-P1A 72

074-579-P1A 74 5. ROUTINE MAINTENANCE The following section outlines the procedures for proper maintenance of the MicroFID II. Maintenance Interval

Seite 73 - 074-579-P1A 73

074-579-P1A 75 Maintenance of the Flame Ionization Detector The MicroFID II detector is a not a field replaceable unit (FRU) and should not be servi

Seite 74 - 5. ROUTINE MAINTENANCE

074-579-P1A 76 3. Replace the filter housing. 4. Calibrate all Cal Memories that you are using before continuing operation. NOTE: Do not oper

Seite 75 - 074-579-P1A 75

074-579-P1A 77 Yearly Service Evaluation INFICON advises that the MicroFID II be sent to INFICON Service once a year for preventive maintenance and

Seite 76 - 074-579-P1A 76

074-579-P1A 78 7. TROUBLE SHOOTING General Information If you have a service related question about the MicroFID II, consult this manual first. If y

Seite 77

074-579-P1A 79 Flame out may also occur when sampling enclosed or confined spaces where vapors and gases cannot escape. Watch for

Seite 78 - 7. TROUBLE SHOOTING

074-579-P1A 8 Notices and Warnings FCC Warning This equipment has been tested and found to comply with the limits for a Class B Digital Device, purs

Seite 79

074-579-P1A 80 Cause: Ambient air is contaminated. Action: If you are unsure about the quality of the ambient air, use a charcoal filter

Seite 80 - MicroFID II Troubleshooting

074-579-P1A 81 Problem: Date and time settings are not retained. Cause: The MicroFID II has not been used for 3 months or more and the internal batt

Seite 81

074-579-P1A 82 Problem: Sample flow rate varies from 600 ml/min. +/- 10%. Cause: Inlet filter has not been installed. Action: Install an inlet

Seite 82

074-579-P1A 83 If you will be using the MicroFID II in a highly contaminated area where it is possible the oxygen content is below 10%, w

Seite 83

074-579-P1A 84 8. APPENDICES MicroFID II Specifications OperatingTemperature Range ... 32°F (0°C) to 122°F (50°C) Operating Humidit

Seite 84 - 8. APPENDICES

074-579-P1A 85 Monitoring Software Monitoring software compatibility: ... Communicates wirelessly with INFICON’s FirstFID software system.

Seite 85 - 074-579-P1A 85

074-579-P1A 86 MicroFID II Telescoping Probe Holder Installation Guide 1. Position the MicroFID II so that the back of the instrument is facing you

Seite 86 - 074-579-P1A 86

074-579-P1A 87 MicroFID II Shoulder Strap Adaptor Kit Installation Guide 1. Kit Includes: a. One Shoulder Strap b. Two Clip Adapters c. Four Sc

Seite 87

074-579-P1A 88 Warranty The MicroFID II is warranted for one year against defects in materials and workmanship. INFICON warrants that its manufactur

Seite 88 - Warranty

074-579-P1A 89 Contacting INFICON To place an order, check the status of an order, obtain current pricing and availability Service and repair of yo

Seite 89 - 315-434-1100

074-579-P1A 9 WARNING If there is any indication of physical damage to the MicroFID II after it has been dropped, the unit must be checked out by