Sensistor ISH2000HYDROGEN LEAK DETECTORPublication: INFICON AB - nipa60e1-a (1202) - All information can be modified without prior noticeTechnical Ref
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 10A 600A 401 About HydrogenWhen pure hydrogen gas is released in air its flammability range spa
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 100Dismantling probe1. Loosen the LEMO contact nut using a spanner for noched probe nuts.Figure
EN- 101 Sensistor ISH2000 Technical Reference Manual (TRM)4. When changing the PCB the cables must be removed by soldering.Figure E-30: 5. To replace
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 102FF 100 ServiceF 800 ToolsAppendixSensistor ISH2000 Technical Reference ManualDetailed conten
EN- 103 Sensistor ISH2000 Technical Reference Manual (TRM)F 100F 100 Support by INFICONHow To Contact INFICONFor Sales and Customer Service contact ne
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 104F 110F 110 Disposal of product when taken out of serviceRoHS and WEEEFor information regardi
EN- 105 Sensistor ISH2000 Technical Reference Manual (TRM)F 800F 800 ToolsSensistor ISH2000 softwareSensistor ISH2000 Installer “Sensistor ISH2000 Ins
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 106F 8006. Turn on the instrument7. Wait for download8. The following windows on next page appe
EN- 107 Sensistor ISH2000 Technical Reference Manual (TRM)F 800APC InstallerInstallation instructions:1. Turn off the instrument2. Connect a a standar
INFICON AB, Box 76, SE-581 02 Linköping, Sweden Phone: +46 (0) 13 35 59 00 Fax: +46 (0) 13 35 59 01 www.inficon.com E-mail: reach.sweden@infi
EN- 11 Sensistor ISH2000 Technical Reference Manual (TRM)A 700A 600 OptionsWhich options for which model?Sensistor ISH2000Sensistor ISH2000 CSensistor
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 12A 700A 700 Sensistor ISH2000 ComponentsSensistor ISH2000Figure A-4: Sensistor ISH2000 compone
EN- 13 Sensistor ISH2000 Technical Reference Manual (TRM)A 700Sensistor ISH2000 components according to Figure A-4: on page 12.Position P/N Position P
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 14A 700Sensistor ISH2000CFigure A-5: Sensistor ISH2000C componentsRed diodeGreen diodeRed diode
EN- 15 Sensistor ISH2000 Technical Reference Manual (TRM)A 700Sensistor ISH2000C components according to Figure A-5: on page 14.Position P/N Position
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 16A 700Sensistor ISH2000PFigure A-6: Sensistor ISH2000P components'ZPE/KZP
EN- 17 Sensistor ISH2000 Technical Reference Manual (TRM)A 701Sensistor ISH2000P components according to Figure A-6: on page 16.Position P/N Position
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 18A 701A 701 Spare parts and accessoriesAP29ECO Sampling ProbeAP55 Sniffer ProbeAP57 Counter Fl
EN- 19 Sensistor ISH2000 Technical Reference Manual (TRM)A 701P50 Hand ProbesH65 Insert SensorProbe Tip Protection Cap and filterHand Probe P50 with r
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 2A 100 Introduction to the Sensistor ISH2000 series ...
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 20A 701ComboxC 21 Probe CablesProbe adapter for AP29ECO, H65, AP55 and AP57.Part No: 590-820In
EN- 21 Sensistor ISH2000 Technical Reference Manual (TRM)A 701Reference LeaksPower cablesFuseLarge and small, with certificate, for calibration and fu
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 22A 800Carrying caseBattery chargerHand Probe sensorMounting kitPhoenix connector to Sensistor
EN- 23 Sensistor ISH2000 Technical Reference Manual (TRM)A 800A 800 Sensistor ISH2000 - Technical specificationsPower supply specificationsInput and o
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 24A 800Miscellaneous specificationsGas sensing specification (In Detection Mode)Short circuit p
EN- 25 Sensistor ISH2000 Technical Reference Manual (TRM)A 900Gas sensing specification (In Analysis Mode)Selected unit Sensitivity MeasurementrangeLi
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 26A 900A 900 DimensionsSensistor ISH2000Figure A-7: Sensistor ISH2000 dimensions275 / 0.9165 /
EN- 27 Sensistor ISH2000 Technical Reference Manual (TRM)Sensistor ISH2000CFigure A-8: Sensistor ISH2000C dimensionsSensistor ISH2000PFigure A-9: Sens
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 28B 100 SafetyB 110 Storage - transportationB 210 Connecting the detector to the installationB
EN- 29 Sensistor ISH2000 Technical Reference Manual (TRM)B 100B 100 Safety instructionsThe normal risks associated with working with all compressed ga
EN- 3 Sensistor ISH2000 Technical Reference Manual (TRM)
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 30B 110B 110 Supplies and storageSuppliesThe following parts are supplied with your detector:St
EN- 31 Sensistor ISH2000 Technical Reference Manual (TRM)B 210 Connecting the detector to the installationPorts and connections Sensistor ISH2000Figur
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 32B 210Sensistor ISH2000C connectionsFigure B-3: Sensistor ISH200 C connectionsSensistor ISH200
EN- 33 Sensistor ISH2000 Technical Reference Manual (TRM)B 300B 300 Controlling the detector with the I/O interfaceI/O InterfaceThe Sensistor ISH2000
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 34B 300Probe Control Port ConnectorThe control port connector is a 25-pin female D-sub.Pin Type
EN- 35 Sensistor ISH2000 Technical Reference Manual (TRM)B 300Status signal patternsStatus signals for pin 14 - 18.Figure B-5: Example screenWhere:DET
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 36B 300Status signal after power onFigure B-6: Status signals after power onStatus signal when
EN- 37 Sensistor ISH2000 Technical Reference Manual (TRM)B 310 Controlling the detector with a PC computer through the RS 232 interfaceRS 232 interfac
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 38B 310Q = Set APC in stand by, (stop a measurement)Returns a “Q”.R = Activates probe 0 (built
EN- 39 Sensistor ISH2000 Technical Reference Manual (TRM)B 310Parameters can be sent in any order you like:• If your data was received and correct, Se
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 4A 100 Introduction to the Sensistor ISH2000 seriesA 300 Sensor Technologies working principleA
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 40B 320 Connecting the detector directly to a printer or another deviceFor detailed information
EN- 41 Sensistor ISH2000 Technical Reference Manual (TRM)B 320Connecting the detector to a computerFor example, the user can communicate easily with t
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 42Probe type determines information printed.Analysis data outputThe Analysis Data Output option
EN- 43 Sensistor ISH2000 Technical Reference Manual (TRM)Example: 2.5E-04R (LF)This example is a line feed (LF), R means that the test was above the R
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 44C 100 Getting started with the leak detectorC 110 Operating principle of the control panelC 2
EN- 45 Sensistor ISH2000 Technical Reference Manual (TRM)C 100C 100 Getting started with the leak detectorStarting up / switching off the leak detecto
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 46C 200C 200 Control panelDescriptionFigure C-1: Sensistor ISH2000 control panelGraphic interfa
EN- 47 Sensistor ISH2000 Technical Reference Manual (TRM)C 200Push-buttonsThe functions of the push-buttons are shown at the lower edge of the display
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 48C 201C 201 Menu systemThe menu system is designed as a tree structure similar to that used in
EN- 49 Sensistor ISH2000 Technical Reference Manual (TRM)C 201Sensistor ISH2000 v. 6.0 Menu SystemFigure C-3: Menu systemMeasure ScreenChange Test Mod
EN- 5 Sensistor ISH2000 Technical Reference Manual (TRM)A 100A 100 Introduction to the Sensistor ISH2000 seriesSensistor ISH2000 can be purchased in o
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 50C 201Change test modeChoose the measuring method you will use in the menu Change Test Mode. T
EN- 51 Sensistor ISH2000 Technical Reference Manual (TRM)C 201APC settingsAPC is an abbreviation for Active Probe Control. The APC function is for the
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 52C 201Display settingsThis section describes the different display settings of the Sensistor I
EN- 53 Sensistor ISH2000 Technical Reference Manual (TRM)C 201means that no password is needed to modify the critical parameters. The default is no pa
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 54C 210C 210 PasswordChange PasswordThe user password is a text string (max 12 alphanumerical c
EN- 55 Sensistor ISH2000 Technical Reference Manual (TRM)C 211C 211 Operation of the leak detectorTo detect leaksIf all you wish to do is to detect th
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 56C 211When the probe is moved around and over the product, the signal will increase as the pro
EN- 57 Sensistor ISH2000 Technical Reference Manual (TRM)C 300C 300 Calibration of the leak detectorIntroductionThe leak detector is the instrument an
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 58C 300Calibration procedureBefore calibration, the Reference Value in the Calibration Menu mus
EN- 59 Sensistor ISH2000 Technical Reference Manual (TRM)C 300Irregular reference warningThe Detector will warn if the calibration signal is unreasona
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 6A 300the operator’s panel or any other flat surface. Also it operates on +24 VDC. Mounting bra
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 60C 300Note: If calibration fails you can still use the instrument. Last valid calibration para
EN- 61 Sensistor ISH2000 Technical Reference Manual (TRM)C 300Minimum calibration timeThis parameter sets the lowest possible Calibration Time that ca
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 62C 304C 304 Correlation valueIf you want to use another unit you have to put a recalculation n
EN- 63 Sensistor ISH2000 Technical Reference Manual (TRM)C 305C 305 Reference values settingsReference value with reference leakWhen measuring leak fl
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 64C 410C 410 HeadphonesThe operator can connect a headphone to the detector.AccessoriesHeadphon
EN- 65 Sensistor ISH2000 Technical Reference Manual (TRM)C 500C 500 Factory configuration of the leak detector parametersDefault parametersRange and d
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 66Number of significantDigits2/3 2Password Max 12 characters “” = No passwordPassword protected
EN- 67 Sensistor ISH2000 Technical Reference Manual (TRM)C 570C 570 Date - Time - Language - UnitSet DateReal Time Clock date set as YY-MM-DD. Clock r
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 68C 580C 580 Service mode och debug modeService settingsThe detector is equipped with a service
EN- 69 Sensistor ISH2000 Technical Reference Manual (TRM)In Service Mode, the following information accessible:In APC I/O, the following information a
EN- 7 Sensistor ISH2000 Technical Reference Manual (TRM)A 400A 300 Sensor Technologies working principleSensor TechnologyThe Sensistor ISH2000 leak de
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 704 Status of external output 4 (OUT_4)5 Status of external output 5 (OUT_5)6 Status of externa
EN- 71 Sensistor ISH2000 Technical Reference Manual (TRM)Debug ModeDo not enter this mode. Debug mode is for the manufacturer's use only. Enter-i
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 72D D 100 Preventive maintenanceD 200 General troubleshooting guideD 300 Refe
EN- 73 Sensistor ISH2000 Technical Reference Manual (TRM)D 100D 100 Preventive maintenanceAs the Sensistor ISH2000 does not contain any mechanical par
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 74D 200D 200 General troubleshooting guideSymptoms descriptionFault symptoms Fault MeasuresErro
EN- 75 Sensistor ISH2000 Technical Reference Manual (TRM)D 200Short operating time on battery modelOld or broken battery Replace the batteryThe time a
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 76D 300D 300 Reference leak trouble guideIt is very important that we advice the end user to ca
EN- 77 Sensistor ISH2000 Technical Reference Manual (TRM)This page intentionally left blank
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 78E 100 Maintenance operations introductionE 120 Sending the leak detection for reparation to a
EN- 79 Sensistor ISH2000 Technical Reference Manual (TRM)E 100E 100 Maintenance operations introductionElectrostatic discharge (ESD) protected equipme
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 8A 400 Testing methodsTo Detect leaksIf all you wish to do is to detect the presence of a leak,
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 80E 120E 120 Sending the leak detection for repair to a service centerProduct return formSee “H
EN- 81 Sensistor ISH2000 Technical Reference Manual (TRM)E 130E 130 Functional verificationSensitivity TestSensitivity Test has been designed to help
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 82E 130Figure E-2: Connect a hose to the gas regulatord. Purge the hose with low flow of refere
EN- 83 Sensistor ISH2000 Technical Reference Manual (TRM)E 130h. Acceptance level for new sensor is minimum 1.2 = 1.20E+00.i. If the Signal is below 1
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 84E 140E 140 Service menuThe Service Mode is reached by starting the instrument and at the same
EN- 85 Sensistor ISH2000 Technical Reference Manual (TRM)E 160E 160 Software and documentation evolutionInstrument software VersionProbe Software Ver-
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 86E 230E 230 Sensistor ISH2000 Main boardComponent locations on boardFigure E-5: Main board
EN- 87 Sensistor ISH2000 Technical Reference Manual (TRM)E 240E 240 Wiring diagram Sensistor ISH2000Main PCBE-6: Main PCBJ7J12J4J8J6J2J14J5J13J9KB 101
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 88E 410E 410 Instrument maintenanceTools neededP/N Sensistor ISH2000 Tools Supplier591-533 Torx
EN- 89 Sensistor ISH2000 Technical Reference Manual (TRM)E 410Dismantling Sensistor ISH20001. Turn off the instrument and remove power supply.2. Unscr
EN- 9 Sensistor ISH2000 Technical Reference Manual (TRM)A 401Hint: It is good practice to detect a leak, locate it, and immediately remove the probe t
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 90E 410Dismantling Sensistor ISH2000C4. Unscrew the four torx screws and remove the power suppl
EN- 91 Sensistor ISH2000 Technical Reference Manual (TRM)E 4102. Open the instrument by pulling out the front as shown in the picture to the right.
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 923. Pull out the front enough toloosen the three connectors.Figure E-11: Pull out the frontFig
EN- 93 Sensistor ISH2000 Technical Reference Manual (TRM)4. Remove the battery as shown in the figures to the right.Figure E-13: Remove screwsFigure E
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 94Dismantling Sensistor ISH2000P1. Turn off the instrument and remove power supply.2. Unscrew t
EN- 95 Sensistor ISH2000 Technical Reference Manual (TRM)Replacement of main PCB1. Separate the cables as shown in the picture to the right.Figure E-1
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 96E 4102. Release the PCB by losing the locknuts on the D-Sub connectors and the LEMO connector
EN- 97 Sensistor ISH2000 Technical Reference Manual (TRM)E 410Removal of speakerRemoval of displayUnscrew the speaker Philip Head Screws as shown in t
Sensistor ISH2000 Technical Reference Manual (TRM) EN- 98E 410Removal of panel buttonsUnscrew the panel button card Philip Head Screws as shown in the
EN- 99 Sensistor ISH2000 Technical Reference Manual (TRM)E 411E 411 Probe maintenanceTools neededChanging the Probe TipP/N P50/P50 Flex Tools Supplier