INTERFACE DESCRIPTIONjins80e1-d (1011)Catalog No.550-300550-310550-330from software version V 1.61Modul1000Helium Leak Detector
1-8 ASCII Protocoljins80e.fm Operating Manual(1011):EXTVAC external test leak in vacuum mode:EXTSNIFF external test leak in sniff mode:CALREQ calibrat
ASCII Protocol 1-9jins80e.fm Operating Manual(1011):LCDDECades number of display decades:LCDInvert invert display:LCDSCALELog display scale lin. / log
1-10 ASCII Protocoljins80e.fm Operating Manual(1011):PLCINLINK :3, 4, 5, 6, 7, 8, 9, 10configuration PLC inputNOt_used, START, STOP, START/STOP, VENT,
ASCII Protocol 1-11jins80e.fm Operating Manual(1011);PURGESETTLINGTimetime from closing the purge valve to opening the valve V2 (2 ... 20s):RECorder r
1-12 ASCII Protocoljins80e.fm Operating Manual(1011):ZERO zero (OFF, ON, STABLE):PEXT1:FULLPpressure when external pressure sensor is rejecting fully:
ASCII Protocol 1-13jins80e.fm Operating Manual(1011)3.5 CalibrationTwo different kinds of calibration can be called via „*CAL“:3.5.1 Internal Automati
1-14 Binary Protocoljins80e.fm Operating Manual(1011)4 Binary Protocol4.1 Command StructureIn binary protocol the command to the leak detector always
Binary Protocol 1-15jins80e.fm Operating Manual(1011)Example 1: Get trigger level 2 in mbarl/sPC Î leak detectorleak detector Î PC:4.3 Data Format56 5
1-16 Binary Protocoljins80e.fm Operating Manual(1011)4.4 Error Messages4.5 Command List230 RS232Host command currently not allowed (host control)231 R
Binary Protocol 1-17jins80e.fm Operating Manual(1011)9 SetSecInMeas Time, since device is in measurement mode No unsigned integer [s]1011GetValveSetVa
1-2 ContentContent jins80e2.fm INTERFACE DESCRIPTION(1011)Content1 Preface 1-12 Interface Connecting Cable 1-23 ASCII Protocol 1-33.1 Command Forma
1-18 Binary Protocoljins80e.fm Operating Manual(1011)18 SetAutomatik Set automatic On/Off No uchar0=Automatic off; 1=Automatic On2021GetTurboCtrlSetTu
Binary Protocol 1-19jins80e.fm Operating Manual(1011)3233GetRangeNVSetRangeNVGet set post amplifier range No unsigned charauto: 0-0.4; 1-1.6; 2-6.4; 3
1-20 Binary Protocoljins80e.fm Operating Manual(1011)4647GetMSKatSetMSKatGet / set active filament No unsigned char(0=Cathode 1, 1=Cathode 2)4849GetMS
Binary Protocol 1-21jins80e.fm Operating Manual(1011)62 GetErrorCode Get current error code No uchar0=No error63 ClearError Clear error No No6465GetSu
1-22 Binary Protocoljins80e.fm Operating Manual(1011)7677GetSuppressionCommanderSetSuppressionCommanderGet / Set Commander suppression Byte 0:0 - off,
Binary Protocol 1-23jins80e.fm Operating Manual(1011)102 GetUp Get pressure gauge voltage(internal & external)Byte 0:1=p12=p23=pext14=pext2float [
1-24 Binary Protocoljins80e.fm Operating Manual(1011)132133GetPartNrSetPartNrGet / set part number No ulint134135GetTimeBaseAutoScaleSetTimeBaseAutoSc
Trouble Shooting 1-25jins80e.fm Operating Manual(1011)5 Trouble Shooting5.1 CommonError Possible Reason SolutionNo characters are received via the int
1-26 Trouble Shootingjins80e.fm Operating Manual(1011)5.2 ASCII ProtocolError Possible Reason SolutionModul1000 does not reply / Modul1000 replies aft
List of menu parameter and ASCII/Binary protocol 1-27jins80e.fm Operating Manual(1011)6 List of menu parameter and ASCII/Binary protocol Menu Paramete
Preface 1-1jins80e.fm Operating Manual(1011)1PrefaceThis document describes the RS232 interface of the Modul1000 and the "ASCII" and"Bi
1-28 List of menu parameter and ASCII/Binary protocoljins80e.fm Operating Manual(1011)Beep sound Main Menu -> Trigger & Alarms -> Volumeon,o
List of menu parameter and ASCII/Binary protocol 1-29jins80e.fm Operating Manual(1011)Machine factor Main Menu -> Settings -> Vacuum settings -&
1-30 List of menu parameter and ASCII/Binary protocoljins80e.fm Operating Manual(1011)Characteristic external pressure gaugeMain Menu -> Settings -
List of menu parameter and ASCII/Binary protocol 1-31jins80e.fm Operating Manual(1011)PLC input Pin 3…10Main Menu -> Settings -> Interfaces ->
1-32 List of menu parameter and ASCII/Binary protocoljins80e.fm Operating Manual(1011)Recorder output scale Upper limitMain Menu ->Settings ->In
List of menu parameter and ASCII/Binary protocol 1-33jins80e.fm Operating Manual(1011)Contamination protectionMain Menu -> Monitoring functions -&g
1-34 List of menu parameter and ASCII/Binary protocoljins80e.fm Operating Manual(1011)t_B zero delay Main Menu -> Settings -> Vacuum settings -&
List of menu parameter and ASCII/Binary protocol 1-35jins80e.fm Operating Manual(1011)p_A Gross leak testMain Menu -> Settings -> Vacuum setting
1-36 List of menu parameter and ASCII/Binary protocoljins80e.fm Operating Manual(1011)cathode Main Menu -> Info -> Service -> Switch filament
List of menu parameter and ASCII/Binary protocol 1-37jins80e.fm Operating Manual(1011)
1-2 Interface Connecting Cablejins80e.fm Operating Manual(1011)2 Interface Connecting CableThe interface (RS232) is wired as data communication equipm
INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, GermanyPhone: +49 (0)221 347-40 Fax: +49 (0)221 347-41429 E-mail: [email protected] S
ASCII Protocol 1-3jins80e.fm Operating Manual(1011)3 ASCII Protocol3.1 Command Format In ASCII protocol any command starts with « * (ASCII code 42dec
1-4 ASCII Protocoljins80e.fm Operating Manual(1011)3.2 Error Messages3.3 Examples3.4 Command List ASCII ModeOK ERR_OK command completedE01 ERR_CMD_STA
ASCII Protocol 1-5jins80e.fm Operating Manual(1011)*STATusstatus of Modul1000 (INIT, ACCL, STBY, VENT, WAIT_EVAC, EVAC, MEAS, CAL, ERROR)[:CAL]mode ca
1-6 ASCII Protocoljins80e.fm Operating Manual(1011)*STOp stop*VENt vent*CAL in STBY → internal automatic calibrationin MEAS → extern manual calibrat
ASCII Protocol 1-7jins80e.fm Operating Manual(1011):IFilter current filtered [A]:LRMAXmax. leak rate since last inquiry via interface in chosen unit[: