kins22e1-j, 1401Interface DescriptionEcotec E3000Leak DetectorOrder No.:530-001, 530-002530-103 ... 530-106from software version:2.61
10 2 I/O PortInterface Description Ecotec E3000, kins22e1-j, 1401Technical Data relay outputsRelay outputActive = Normally open contact (NO) closed
2 I/O Port 11Interface Description Ecotec E3000, kins22e1-j, 14011.0 … < 3 V: 1st decade3.0 … < 5 V: 2nd decade5.0 … < 7 V: 3rd decade7.
12 2 I/O PortInterface Description Ecotec E3000, kins22e1-j, 1401Recorder output gas Here one gas from the four gases is defined which will be outpu
2 I/O Port 13Interface Description Ecotec E3000, kins22e1-j, 1401To change these settings select the appropriate pin with the UP and DOWN arrows on
14 2 I/O PortInterface Description Ecotec E3000, kins22e1-j, 1401Fig. 5: Calibration timing intervalA calibration process is started by setting the
3 RS-232 interface 15Interface Description Ecotec E3000, kins22e1-j, 14013 RS-232 interfaceThe Ecotec E3000 is equipped with a RS-232 interface whi
16 3 RS-232 interfaceInterface Description Ecotec E3000, kins22e1-j, 14013.2 Interface Commands3.2.1 Command Format In ASCII mode any command star
3 RS-232 interface 17Interface Description Ecotec E3000, kins22e1-j, 14013.2.2 List of CommandsThe following table lists all commands available in
18 3 RS-232 interfaceInterface Description Ecotec E3000, kins22e1-j, 1401:ECOcheck ECO-Check (ON, OFF) R/S:ENDsign end sign (CR, LF, CR LF) R/S:FILa
3 RS-232 interface 19Interface Description Ecotec E3000, kins22e1-j, 1401:SELect gas select (Auto, Manu, Hold) R/S:SELTime Gas select time (5, 10, 1
2 Interface Description Ecotec E3000, 1401Reprint, translation and reproduction require the written permission of the INFICON GmbH.
20 3 RS-232 interfaceInterface Description Ecotec E3000, kins22e1-j, 1401:MODE Modeenable: on, 1, ENAbledisable: off, 0, DISableR/S:NAME Gas name
3 RS-232 interface 21Interface Description Ecotec E3000, kins22e1-j, 1401:TURBO operating hours turbo pump [h] R*MEASure R:Argon R:Current current f
22 3 RS-232 interfaceInterface Description Ecotec E3000, kins22e1-j, 1401:NAME name (max. 6 characters) R/S:NR:NRAgas number (1 … 4) R/S:NRB Gas nu
3 RS-232 interface 23Interface Description Ecotec E3000, kins22e1-j, 1401[:CAL] status calibration (status number)2 - LEAKRATE STABLE, CONFIRM4- B
24 3 RS-232 interfaceInterface Description Ecotec E3000, kins22e1-j, 1401[:SNkey] State Sniffer keyByte 1 leftByte 2 rightR[:LEAK] ECO-Check t
3 RS-232 interface 25Interface Description Ecotec E3000, kins22e1-j, 1401[:WARNing] State warning flagsByte 1 – CapillaryByte 2 – CalRequestByte
26 3 RS-232 interfaceInterface Description Ecotec E3000, kins22e1-j, 14013.2.3 Error MessagesThe following error messages may be received from the
3 RS-232 interface 27Interface Description Ecotec E3000, kins22e1-j, 1401Diagnostics The diagnostics protocol is a binary protocol used for analysis
28 3 RS-232 interfaceInterface Description Ecotec E3000, kins22e1-j, 14013.4 Examples of programming3.4.1 Operating in measurement modeGas setting
3 RS-232 interface 29Interface Description Ecotec E3000, kins22e1-j, 14013.4.2 Performing an external calibrationGas settings and interface setting
Table of Contents 3Interface Description Ecotec E3000, kins22e1-j, 1401Table of Contents1 About this manual ...
30 3 RS-232 interfaceInterface Description Ecotec E3000, kins22e1-j, 1401*cal:status? (CR) Requests the status of the calibration.LEAK STABLE, CONFI
3 RS-232 interface 31Interface Description Ecotec E3000, kins22e1-j, 1401CAL (CR) Status: still in calibration mode (calibration process not finishe
32 3 RS-232 interfaceInterface Description Ecotec E3000, kins22e1-j, 1401
4 Trouble Shooting 33Interface Description Ecotec E3000, kins22e1-j, 14014 Trouble ShootingTable 15: Trouble Shooting Error Possible Reason Solutio
34 4 Trouble ShootingInterface Description Ecotec E3000, kins22e1-j, 1401
35Interface Description Ecotec E3000, kins22e1-j, 1401
INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, GermanyPhone: +49 (0)221 347-40Fax: +49 (0)221 347-41429E-mail: [email protected]UNITED S
4 Table of Contents Interface Description Ecotec E3000, kins22e1-j, 1401
1 About this manual 5Interface Description Ecotec E3000, kins22e1-j, 14011 About this manual1.1 Target groupsThis Interface Description is intende
6 1 About this manualInterface Description Ecotec E3000, kins22e1-j, 1401
2 I/O Port 7Interface Description Ecotec E3000, kins22e1-j, 14012 I/O PortThe I/O part allows communication with external equipment via analog data
8 2 I/O PortInterface Description Ecotec E3000, kins22e1-j, 14012.1.1 PLC inputsTechnical data PLC inputs 24 V nominal input.– Low level: 0 … 7 V–
2 I/O Port 9Interface Description Ecotec E3000, kins22e1-j, 1401Gas select, gas a, gas b Through the control inputs Gas … (pin 9, 13 and 25) it is p
Kommentare zu diesen Handbüchern