DataFID™Flame Ionization DetectorIPN 074-578-P1AOPERATING MANUAL
074-578-P1A 10 WARNING 1. All calibration, maintenance and servicing of this device, including battery charging, must be performed in a safe are
074-578-P1A 11 DECLARATION OF CONFORMITY This is to certify that this equipment, designed and manufactured by: INFICON Inc. Two Technology Plac
074-578-P1A 12 RoHS: INFICON hereby declares that the above mentioned products are assigned to the classification of Category 9, Industri
074-578-P1A 13 1. INTRODUCTION About this Manual This manual provides detailed instructions for the setup, operation and maintenance of the Data
074-578-P1A 14 UNPACKING THE DataFID The following accessories are included with your DataFID: DataFID Instrument CD Manual 70 Liter hydrogen
074-578-P1A 15 Flame Ionization Detector Operation The DataFID uses a flame ionization detector for the measurement of combustible organic compo
074-578-P1A 16 Figure 2. Block Diagram Detector Response The DataFID is strictly an organic compound detector. It does not respond to inorgani
074-578-P1A 17 Support Equipment and Consumables Compressed Gases Cylinders of compressed gas, such as hydrogen and calibration gas, must be han
074-578-P1A 18 Hydrogen Gas The DataFID detector uses a hydrogen flame to ionize samples and produce Total Volatile Organic Compound (TVOC) rea
074-578-P1A 19 Flammable Gases High concentrations of flammable gases (gases within their flammable range) can act as an additional fuel source.
074-578-P1A 2 Warning: Limitation of Liability The ultimate responsibility of the consequences of use of toxic compounds rests with the user. I
074-578-P1A 20 To connect a charcoal filter 1. Load the Teflon ferrules into the nut. The nut and ferrules are supplied with the filter. See Fig
074-578-P1A 21 2. USING THE DataFID DataFID Feature Layout The photograph below shows the basic layout of the DataFID components. The DataFID c
074-578-P1A 22 LDAR Probe and PDA Figure 6. LDARProbe and PDA WARNING The PDA must be rated intrinsically safe in order to operate in a haza
074-578-P1A 23 Using the DataFID Battery Charger READ THESE INSTRUCTIONS BEFORE USING THE CHARGER The DataFID battery charger is only designed f
074-578-P1A 24 CHARGE CYCLE AND LED INDICATIONS LED MODE Orange Battery not connected Orange Battery initialization & analysis Red Fast Cha
074-578-P1A 25 To Charge the Battery 1. To charge the DataFID battery, unscrew the cover for the charging connector cap. 2. Connect the char
074-578-P1A 26 NOTE: The DataFID will not operate without the charging connector cap in place. NOTE: The DataFID will not operate with the AC A
074-578-P1A 27 WARNING Hydrogen storage on the metal oxide is an Exothermic Reaction. The cylinder may become hot during the initial fill. Car
074-578-P1A 28 Table 2. Charging Time and Usage Time WARNING Use only the DataFID Hydrogen Fill Station p/n A1201222 to fill the hydrogen fue
074-578-P1A 29 CAUTION: When filling the DataFID hydrogen fuel cylinder, ensure that the hydrogen fuel cylinder on/off valve is completely open
074-578-P1A 3 Release History Part Number Release Publication Date MN201101-01 A Dec. 23, 2009 B July 20, 2011 C Sept. 19, 2011 D July 31,
074-578-P1A 30 NOTE: Read through this section before filling the hydrogen fuel cylinder. Store the hydrogen supply tank in a well-ventilated a
074-578-P1A 31 Filling the Hydrogen Fuel Cylinder 1. Rotate the output valve to point toward the main cylinder of hydrogen (Filling Position)
074-578-P1A 32 Figure 10. Hydrogen Fuel Cylinder Connected to Filling Station Hydrogen Tank Valve Warning: Do not adjus
074-578-P1A 33 Installing the Hydrogen Fuel Cylinder in the DataFID 1. If the hydrogen fuel cylinder retention plate is installed on the DataF
074-578-P1A 34 I o WARNING The hydrogen fuel cylinder retention plate MUST be installed on the DataFID to retain the intrinsic safety rating.
074-578-P1A 35 8. The DataFID will automatically start with “DataFID Setup” on the display. Simply press the middle red soft key (“meas”) to ini
074-578-P1A 36 Response Factors for Gases and Vapors To use the response factors 1. Press the CAL key and enter the response factor for the s
074-578-P1A 37 If you will be in the field for more than one day you should include the following additional items:
074-578-P1A 38 3. USER FUNCTIONS Display The DataFID has a graphic display for reporting detected concentrations and to guide the user through
074-578-P1A 39 Menu Function Displays Figure 12:. The DataFID Main Menu Display The display reports instantaneous concentrations at all
074-578-P1A 4 Table of Contents For Your Safety ...
074-578-P1A 40 DataFID Setup Menu Audio ON Audio OFF Pump ON Pump OFF Logging OFF Tag Logging Interval Logging Method 21 Log Optional Air ON Op
074-578-P1A 41 Figure 15 . Shut Down DataFID Menu Figure 16. View Menu Shut Down DataFID Shut down in process Press ‘next’ soft key
074-578-P1A 42 Keys Fixed ON/OFF Key The ON/OFF key is used to both power on to the DataFID and to power off the DataFID. To turn on the Data
074-578-P1A 43 enter password Numeric Value, Duration, Time and Date Entry In cases where the system requir
074-578-P1A 44 DataFID Display Description TVOC Display Max Err TVOC concentration over instrument operating range Hi Chamber temp Chamber tempe
074-578-P1A 45 enter password 1/02/09 location
074-578-P1A 46 1/02/09 meas 14:45 Location mode is identified by the 20 characte
074-578-P1A 47 1/02/09 method 21 14:45 1/02/09 method
074-578-P1A 48 1/02/09 method 21 14:45 1/02/09 method 21
074-578-P1A 49 meth 21 status performed, the status key should be selected by pressing the ‘status’
074-578-P1A 5 Preparing for Field Operation ... 36 Op
074-578-P1A 50 enter password Interval Mode – Data Logging Interval mode logs readings at user selected in
074-578-P1A 51 The second option is the “Download to PC” which begins the download of stored data from the DataFID to the PC. The Bluetooth conn
074-578-P1A 52 2) Example Method 21 File meth21 1 "Method 21 Location 1" "First Meth21 Drawing" "AAA" 10.0 meth21
074-578-P1A 53 Method 21 routing information can be added or subtracted from the file as necessary. Upload Location Route Entry from PC 1) S
074-578-P1A 54 2) Example location route file locroute 1 "First Route Location " locroute 2 "Second Route Location" locrout
074-578-P1A 55
074-578-P1A 56 Upload Compound Entry from PC 1) Single Compound Entry Description A single Compound list entry will take place in the following
074-578-P1A 57 4) Viewing the compound list in the DataFID In order to verify that the Compound list was successfully transferred to the DataFI
074-578-P1A 58 DataFID Setup Functions DataFID Setup functions are used to select the DataFID features (See Figure 14). There are fourteen funct
074-578-P1A 59 enter password Logging Off mode displays instantaneous readings only. The display is
074-578-P1A 6 Connecting the DataFID with Gas Bag and the Zero Air Bag ... 66 General Procedure to Cali
074-578-P1A 60 The up and down arrow soft keys are used to change the character underlined by the cursor. The right arrow is used to advance t
074-578-P1A 61 memory slot 1 Real Time Data When the DataFID is being used in conjunction with a Computer or
074-578-P1A 62 Edit H2 Timer The H2 Timer is used to count down the number of hours of use remaining in the hydrogen fuel cylinder. Each time th
074-578-P1A 63 DataFID Calibration Calibration Technical Description The DataFID will always detect all ionizable compounds present in a sample
074-578-P1A 64 Methane in air is recommended as span gas. However, Hexane or Propane can also be used as span gas. The choice is made from the
074-578-P1A 65 6. Open the regulator by turning the valve counter clockwise, until the ball is approximately 1/8” (3 mm) from its rest position
074-578-P1A 66 Connecting the DataFID with the Calibration Gas Bag and the Zero Air Bag Use the calibration kit (Part No. MX396011) as follows:
074-578-P1A 67 8. The union should be connected to the gas bag adapter. If it is not, then tighten the nut on the adapter tube to the union. 9.
074-578-P1A 68 select span gas connect zero General Procedure to Calibrate the DataFID B
074-578-P1A 69 zero cal zero too high span level Should
074-578-P1A 7 For Your Safety Strictly follow the Instructions for Use Any use of this instrument requires a full understanding and strict adh
074-578-P1A 70 connect span calibrate span edit span until the cur
074-578-P1A 71 span error span completed Should a fault condition arise such as “Concentr
074-578-P1A 72 Multi-Point Calibration In multi-point calibration mode, the DataFID can calibrate up to 5 leak definitions, plus zero. The DataF
074-578-P1A 73 Connect the DataFID to the zero air supply and open the valve to the zero air supply. Then, press the Next key (right hand red ke
074-578-P1A 74 9. The DataFID will automatically enter the span value. If there is a problem with the span gas process, the DataFID will inform
074-578-P1A 75 13. The DataFID will automatically enter span gas 2. 14. Once the last calibration step is performed, the display will show:
074-578-P1A 76 4. WIRELESS COMMUNICATION The DataFID uses a Bluetooth communication chip as the means to download and upload information to and
074-578-P1A 77 Enabling Blue Tooth Communication Users of Computer/PDA datalogging software programs will need to perform the following steps i
074-578-P1A 78 4. Select Real Time Data output ON, then select the Back option 2 times to enter the top level menu or select Measure to resume
074-578-P1A 79 5. ROUTINE MAINTENANCE The following section outlines the procedures for proper maintenance of the DataFID Maintenance Intervals
074-578-P1A 8 Notices and Warnings FCC Warning This equipment has been tested and found to comply with the limits for a Class B Digital Device,
074-578-P1A 80 Maintenance of the Flame Ionization Detector The DataFID detector is a not a field replaceable unit therefore should not be servi
074-578-P1A 81 3. Replace the filter housing. 4. Calibrate all Cal Memories that you are using before continuing operation. NOTE: Do
074-578-P1A 82 once a month and recorded to verify if any changes in the quality check have changed and if the unit should be sent in for servic
074-578-P1A 83 6. TROUBLE SHOOTING General Information This chapter provides cause and action on commonly found faults on the DataFID instrume
074-578-P1A 84 Flame out may also occur when sampling enclosed or confined spaces where vapors and gases cannot escape. Watch for
074-578-P1A 85 Cause: Ambient air is contaminated. Action: If you are unsure about the quality of the ambient air, use a charcoal filter
074-578-P1A 86 Problem: Date and time settings are not retained. Cause: The DataFID has not been used for 3 months or more and the internal batt
074-578-P1A 87 Problem: Sample flow rate varies from 600 ml/min. +/- 10%. Cause: Inlet filter has not been installed. Action: Install an i
074-578-P1A 88 If you will be using the DataFID in a highly contaminated area where it is possible the oxygen content is below 10%, watch for in
074-578-P1A 89 7. APPENDICES DataFID Specifications EPA Standards Meets and exceeds US EPA Method 21 regulations for fugitive emissions monito
074-578-P1A 9 WARNING If there is any indication of physical damage to the DataFID after it has been dropped, the unit must be checked out by
074-578-P1A 90 Warranty The DataFID is warranted for one year against defects in materials and workmanship. INFICON, Inc. warrants that its manu
074-578-P1A 91 Contacting INFICON To place an order, check the status of an order, obtain current pricing and availability Service and repair o