DataFID
Flame Ionization Detector
IPN 074-578-P1A
OPERATING MANUAL
Seitenansicht 0
1 2 3 4 5 6 ... 90 91

Inhaltsverzeichnis

Seite 1 - Flame Ionization Detector

DataFID™Flame Ionization DetectorIPN 074-578-P1AOPERATING MANUAL

Seite 2

074-578-P1A 10 WARNING 1. All calibration, maintenance and servicing of this device, including battery charging, must be performed in a safe are

Seite 3 - Notices

074-578-P1A 11 DECLARATION OF CONFORMITY This is to certify that this equipment, designed and manufactured by: INFICON Inc. Two Technology Plac

Seite 4 - Table of Contents

074-578-P1A 12 RoHS: INFICON hereby declares that the above mentioned products are assigned to the classification of Category 9, Industri

Seite 5

074-578-P1A 13 1. INTRODUCTION About this Manual This manual provides detailed instructions for the setup, operation and maintenance of the Data

Seite 6 - 074-578-P1A 6

074-578-P1A 14 UNPACKING THE DataFID The following accessories are included with your DataFID: DataFID Instrument CD Manual 70 Liter hydrogen

Seite 7 - For Your Safety

074-578-P1A 15 Flame Ionization Detector Operation The DataFID uses a flame ionization detector for the measurement of combustible organic compo

Seite 8 - Notices and Warnings

074-578-P1A 16 Figure 2. Block Diagram Detector Response The DataFID is strictly an organic compound detector. It does not respond to inorgani

Seite 9

074-578-P1A 17 Support Equipment and Consumables Compressed Gases Cylinders of compressed gas, such as hydrogen and calibration gas, must be han

Seite 10 - WARNING

074-578-P1A 18 Hydrogen Gas The DataFID detector uses a hydrogen flame to ionize samples and produce Total Volatile Organic Compound (TVOC) rea

Seite 11

074-578-P1A 19 Flammable Gases High concentrations of flammable gases (gases within their flammable range) can act as an additional fuel source.

Seite 12

074-578-P1A 2 Warning: Limitation of Liability The ultimate responsibility of the consequences of use of toxic compounds rests with the user. I

Seite 13 - 1. INTRODUCTION

074-578-P1A 20 To connect a charcoal filter 1. Load the Teflon ferrules into the nut. The nut and ferrules are supplied with the filter. See Fig

Seite 14 - UNPACKING THE DataFID

074-578-P1A 21 2. USING THE DataFID DataFID Feature Layout The photograph below shows the basic layout of the DataFID components. The DataFID c

Seite 15

074-578-P1A 22 LDAR Probe and PDA Figure 6. LDARProbe and PDA WARNING The PDA must be rated intrinsically safe in order to operate in a haza

Seite 16 - Detector

074-578-P1A 23 Using the DataFID Battery Charger READ THESE INSTRUCTIONS BEFORE USING THE CHARGER The DataFID battery charger is only designed f

Seite 17

074-578-P1A 24 CHARGE CYCLE AND LED INDICATIONS LED MODE Orange Battery not connected Orange Battery initialization & analysis Red Fast Cha

Seite 18

074-578-P1A 25 To Charge the Battery 1. To charge the DataFID battery, unscrew the cover for the charging connector cap. 2. Connect the char

Seite 19 - ACCESSORIES

074-578-P1A 26 NOTE: The DataFID will not operate without the charging connector cap in place. NOTE: The DataFID will not operate with the AC A

Seite 20 - To DataFID

074-578-P1A 27 WARNING Hydrogen storage on the metal oxide is an Exothermic Reaction. The cylinder may become hot during the initial fill. Car

Seite 21 - 2. USING THE DataFID

074-578-P1A 28 Table 2. Charging Time and Usage Time WARNING Use only the DataFID Hydrogen Fill Station p/n A1201222 to fill the hydrogen fue

Seite 22

074-578-P1A 29 CAUTION: When filling the DataFID hydrogen fuel cylinder, ensure that the hydrogen fuel cylinder on/off valve is completely open

Seite 23

074-578-P1A 3 Release History Part Number Release Publication Date MN201101-01 A Dec. 23, 2009 B July 20, 2011 C Sept. 19, 2011 D July 31,

Seite 24

074-578-P1A 30 NOTE: Read through this section before filling the hydrogen fuel cylinder. Store the hydrogen supply tank in a well-ventilated a

Seite 25

074-578-P1A 31 Filling the Hydrogen Fuel Cylinder 1. Rotate the output valve to point toward the main cylinder of hydrogen (Filling Position)

Seite 26

074-578-P1A 32 Figure 10. Hydrogen Fuel Cylinder Connected to Filling Station Hydrogen Tank Valve Warning: Do not adjus

Seite 27

074-578-P1A 33 Installing the Hydrogen Fuel Cylinder in the DataFID 1. If the hydrogen fuel cylinder retention plate is installed on the DataF

Seite 28

074-578-P1A 34 I o WARNING The hydrogen fuel cylinder retention plate MUST be installed on the DataFID to retain the intrinsic safety rating.

Seite 29 - Hydrogen Cylinder

074-578-P1A 35 8. The DataFID will automatically start with “DataFID Setup” on the display. Simply press the middle red soft key (“meas”) to ini

Seite 30

074-578-P1A 36 Response Factors for Gases and Vapors To use the response factors 1. Press the CAL key and enter the response factor for the s

Seite 31 - Venting Position

074-578-P1A 37 If you will be in the field for more than one day you should include the following additional items:

Seite 32

074-578-P1A 38 3. USER FUNCTIONS Display The DataFID has a graphic display for reporting detected concentrations and to guide the user through

Seite 33 - 074-578-P1A 33

074-578-P1A 39 Menu Function Displays Figure 12:. The DataFID Main Menu Display The display reports instantaneous concentrations at all

Seite 34

074-578-P1A 4 Table of Contents For Your Safety ...

Seite 35 - Method 21 Operation

074-578-P1A 40 DataFID Setup Menu Audio ON Audio OFF Pump ON Pump OFF Logging OFF Tag Logging Interval Logging Method 21 Log Optional Air ON Op

Seite 36 - Field Check List

074-578-P1A 41 Figure 15 . Shut Down DataFID Menu Figure 16. View Menu Shut Down DataFID Shut down in process Press ‘next’ soft key

Seite 37

074-578-P1A 42 Keys Fixed ON/OFF Key The ON/OFF key is used to both power on to the DataFID and to power off the DataFID. To turn on the Data

Seite 38 - 3. USER FUNCTIONS

074-578-P1A 43 enter password Numeric Value, Duration, Time and Date Entry In cases where the system requir

Seite 39

074-578-P1A 44 DataFID Display Description TVOC Display Max Err TVOC concentration over instrument operating range Hi Chamber temp Chamber tempe

Seite 40 - DataFID Setup Menu

074-578-P1A 45 enter password 1/02/09 location

Seite 41

074-578-P1A 46 1/02/09 meas 14:45 Location mode is identified by the 20 characte

Seite 42 - 074-578-P1A 42

074-578-P1A 47 1/02/09 method 21 14:45 1/02/09 method

Seite 43 - Figure 7: Edit Mode Display

074-578-P1A 48 1/02/09 method 21 14:45 1/02/09 method 21

Seite 44 - Table 5. System Alerts

074-578-P1A 49 meth 21 status performed, the status key should be selected by pressing the ‘status’

Seite 45 - 35.7 ppm

074-578-P1A 5 Preparing for Field Operation ... 36 Op

Seite 46 -

074-578-P1A 50 enter password Interval Mode – Data Logging Interval mode logs readings at user selected in

Seite 47 - 1.72 ppm

074-578-P1A 51 The second option is the “Download to PC” which begins the download of stored data from the DataFID to the PC. The Bluetooth conn

Seite 48 - 100 ppm

074-578-P1A 52 2) Example Method 21 File meth21 1 "Method 21 Location 1" "First Meth21 Drawing" "AAA" 10.0 meth21

Seite 49 - 074-578-P1A 49

074-578-P1A 53 Method 21 routing information can be added or subtracted from the file as necessary. Upload Location Route Entry from PC 1) S

Seite 50 - Enter new INTERVAL (secs)

074-578-P1A 54 2) Example location route file locroute 1 "First Route Location " locroute 2 "Second Route Location" locrout

Seite 52 - 074-578-P1A 52

074-578-P1A 56 Upload Compound Entry from PC 1) Single Compound Entry Description A single Compound list entry will take place in the following

Seite 53 - 074-578-P1A 53

074-578-P1A 57 4) Viewing the compound list in the DataFID In order to verify that the Compound list was successfully transferred to the DataFI

Seite 54 - 074-578-P1A 54

074-578-P1A 58 DataFID Setup Functions DataFID Setup functions are used to select the DataFID features (See Figure 14). There are fourteen funct

Seite 55 - 074-578-P1A 55

074-578-P1A 59 enter password Logging Off mode displays instantaneous readings only. The display is

Seite 56 - 074-578-P1A 56

074-578-P1A 6 Connecting the DataFID with Gas Bag and the Zero Air Bag ... 66 General Procedure to Cali

Seite 57 - 074-578-P1A 57

074-578-P1A 60 The up and down arrow soft keys are used to change the character underlined by the cursor. The right arrow is used to advance t

Seite 58 - 074-578-P1A 58

074-578-P1A 61 memory slot 1 Real Time Data When the DataFID is being used in conjunction with a Computer or

Seite 59 - Enter current time

074-578-P1A 62 Edit H2 Timer The H2 Timer is used to count down the number of hours of use remaining in the hydrogen fuel cylinder. Each time th

Seite 60 - 074-578-P1A 60

074-578-P1A 63 DataFID Calibration Calibration Technical Description The DataFID will always detect all ionizable compounds present in a sample

Seite 61 - 074-578-P1A 61

074-578-P1A 64 Methane in air is recommended as span gas. However, Hexane or Propane can also be used as span gas. The choice is made from the

Seite 62 - 074-578-P1A 62

074-578-P1A 65 6. Open the regulator by turning the valve counter clockwise, until the ball is approximately 1/8” (3 mm) from its rest position

Seite 63 - 074-578-P1A 63

074-578-P1A 66 Connecting the DataFID with the Calibration Gas Bag and the Zero Air Bag Use the calibration kit (Part No. MX396011) as follows:

Seite 64

074-578-P1A 67 8. The union should be connected to the gas bag adapter. If it is not, then tighten the nut on the adapter tube to the union. 9.

Seite 65 - 074-578-P1A 65

074-578-P1A 68 select span gas connect zero General Procedure to Calibrate the DataFID B

Seite 66

074-578-P1A 69 zero cal zero too high span level Should

Seite 67 - 074-578-P1A 67

074-578-P1A 7 For Your Safety Strictly follow the Instructions for Use Any use of this instrument requires a full understanding and strict adh

Seite 68 - 074-578-P1A 68

074-578-P1A 70 connect span calibrate span edit span until the cur

Seite 69 - 074-578-P1A 69

074-578-P1A 71 span error span completed Should a fault condition arise such as “Concentr

Seite 70 - 074-578-P1A 70

074-578-P1A 72 Multi-Point Calibration In multi-point calibration mode, the DataFID can calibrate up to 5 leak definitions, plus zero. The DataF

Seite 71 - 074-578-P1A 71

074-578-P1A 73 Connect the DataFID to the zero air supply and open the valve to the zero air supply. Then, press the Next key (right hand red ke

Seite 72 - 074-578-P1A 72

074-578-P1A 74 9. The DataFID will automatically enter the span value. If there is a problem with the span gas process, the DataFID will inform

Seite 73 - 074-578-P1A 73

074-578-P1A 75 13. The DataFID will automatically enter span gas 2. 14. Once the last calibration step is performed, the display will show:

Seite 74 - 074-578-P1A 74

074-578-P1A 76 4. WIRELESS COMMUNICATION The DataFID uses a Bluetooth communication chip as the means to download and upload information to and

Seite 75 - 074-578-P1A 75

074-578-P1A 77 Enabling Blue Tooth Communication Users of Computer/PDA datalogging software programs will need to perform the following steps i

Seite 76 - 4. WIRELESS COMMUNICATION

074-578-P1A 78 4. Select Real Time Data output ON, then select the Back option 2 times to enter the top level menu or select Measure to resume

Seite 77 - 074-578-P1A 77

074-578-P1A 79 5. ROUTINE MAINTENANCE The following section outlines the procedures for proper maintenance of the DataFID Maintenance Intervals

Seite 78 - 074-578-P1A 78

074-578-P1A 8 Notices and Warnings FCC Warning This equipment has been tested and found to comply with the limits for a Class B Digital Device,

Seite 79 - 5. ROUTINE MAINTENANCE

074-578-P1A 80 Maintenance of the Flame Ionization Detector The DataFID detector is a not a field replaceable unit therefore should not be servi

Seite 80 - 074-578-P1A 80

074-578-P1A 81 3. Replace the filter housing. 4. Calibrate all Cal Memories that you are using before continuing operation. NOTE: Do

Seite 81 - 074-578-P1A 81

074-578-P1A 82 once a month and recorded to verify if any changes in the quality check have changed and if the unit should be sent in for servic

Seite 82

074-578-P1A 83 6. TROUBLE SHOOTING General Information This chapter provides cause and action on commonly found faults on the DataFID instrume

Seite 83 - 6. TROUBLE SHOOTING

074-578-P1A 84 Flame out may also occur when sampling enclosed or confined spaces where vapors and gases cannot escape. Watch for

Seite 84

074-578-P1A 85 Cause: Ambient air is contaminated. Action: If you are unsure about the quality of the ambient air, use a charcoal filter

Seite 85 - DataFID Troubleshooting

074-578-P1A 86 Problem: Date and time settings are not retained. Cause: The DataFID has not been used for 3 months or more and the internal batt

Seite 86

074-578-P1A 87 Problem: Sample flow rate varies from 600 ml/min. +/- 10%. Cause: Inlet filter has not been installed. Action: Install an i

Seite 87

074-578-P1A 88 If you will be using the DataFID in a highly contaminated area where it is possible the oxygen content is below 10%, watch for in

Seite 88

074-578-P1A 89 7. APPENDICES DataFID Specifications EPA Standards Meets and exceeds US EPA Method 21 regulations for fugitive emissions monito

Seite 89 - 7. APPENDICES

074-578-P1A 9 WARNING If there is any indication of physical damage to the DataFID after it has been dropped, the unit must be checked out by

Seite 90 - Warranty

074-578-P1A 90 Warranty The DataFID is warranted for one year against defects in materials and workmanship. INFICON, Inc. warrants that its manu

Seite 91 - 315-434-1100

074-578-P1A 91 Contacting INFICON To place an order, check the status of an order, obtain current pricing and availability Service and repair o