32 LD Protocol
10 A
TMP nominal
status
R/W UINT8
TMP nominal status
0 = off
1 = on
11 B
Calibration
acknowledge
WUINT8
1 = Continue calibration
0 = cancel calibration
12 C
Open/close int.
testleak
R/W UINT8
0 = close 1 = open
incl. Emission monitoring (less sensitive)
internal calibration will overwrite the state
128 80
Leak rate [sel.
unit]
R FLOAT Leak rate in selected unit
129 81
Leak rate
[mbar*l/s]
R FLOAT Leak rate in mbar*l/s
130 82
Internal
pressure 1 [sel.
unit]
R FLOAT Pressure p1 in selected unit
131 83
Internal
pressure 1
[mbar]
R FLOAT Pressure p1 in mbar
132 84
Internal
Pressure 2 [sel.
unit]
R FLOAT Pressure p2 in selected unit
133 85
Internal
Pressure 2
[mbar]
R FLOAT Pressure p2
134 86
Pressure
sensor 3
RFLOAT
Sensor (0...10 V). Config via commands
2630,2634,2638
135 87
Pressure
sensor 4
RFLOAT
Sensor (0...20 mA) config via commands
2632,2636,2639
138 8A
TMP actual
rotation speed
[Hz]
R UINT16 TMP actual rotation speed
139 8B TMP power [W] R FLOAT
TMP power in Watt as reportet by TMP
controller
140 8C
TMP operation
hours [h]
R UINT32 TMP operation hours
141 8D
Frequency
converter
operation hours
[h]
R UINT32 Frequency converter operation hours [h]
142 8E
Leak detector
operation hours
R UINT32 Leak detector operation hours
143 8F
TMP
temperature
bottom [deg. C]
R FLOAT TMP temperatur bottom [deg. C]
144 90
TMP
temperature
electronic [deg.
C]
R FLOAT TMP temperatur electronic [deg. C]
Comm
and dez
Com
mand
hex
Name R/W Data type
Min-, Def.-, Max-
value
Meaning