tinb14e1 (2009-07) 1 Pirani Gauge Display PGD400 set Operating Manual Incl. EC Declaration of Conformity
10 tinb14e1 (2009-07) Measurement values Measurement range (air, O2, CO, N2) 5×10-4 … 1000 mbar (+1.9 … +10.0 VDC) (→ [1]) Measurement error
tinb14e1 (2009-07) 11 Switching function relay Contact type floating changeover contact Load max. 60 VDC, 1 A (ohmic) 50 VAC, 5 A (ohmic) Ser
12 tinb14e1 (2009-07) 3 Installation DANGER DANGER: damaged product Putting a damaged product into operation can be extremely hazardous. In ca
tinb14e1 (2009-07) 13 For mounting the PGD400 into a control panel, the following cut-out is required: 10591.27869M3 or ø 3.5 The admissible ma
14 tinb14e1 (2009-07) 3.2 Mains Power Connector DANGER DANGER: line voltage Incorrectly grounded products can be extremely hazardous in the ev
tinb14e1 (2009-07) 15 Grounding On the rear of the unit, there is a screw which can be used to connect the unit to ground, e.g. using the groun
16 tinb14e1 (2009-07) Pin assignment SENSOR Pin assignment of the 8-pole RJ45 appli-ance connector: 81 Pin Signal 4 1 2 3 5 6 7
tinb14e1 (2009-07) 17 DANGER DANGER: protective low voltage According to EN 61010, voltages exceeding 30 VAC or 60 VDC are hazardous. Only con
18 tinb14e1 (2009-07) 4 Operation 4.1 Front Panel setOperator keyssetPressureunitSwitching functionstatusThresholdMeasurement value in floatin
tinb14e1 (2009-07) 19 4.2 Turning the PGD400 On and Off Make sure the PGD400 is correctly installed and the specifica-tions in the Technical D
2 tinb14e1 (2009-07) Product Identification In all communications with INFICON, please specify the infor-mation on the product nameplate. For co
20 tinb14e1 (2009-07) 4.3 Operating Modes The PGD400 works in the following operating modes: • Measurement mode for displaying measurement value
tinb14e1 (2009-07) 21 4.4 Measurement Mode The Measurement mode is the standard operating mode of the PGD400. Measurement values as well as sta
22 tinb14e1 (2009-07) Getting to the Parameter mode → 22 4.5 Parameter Mode The Parameter mode is used for displaying, editing and entering
tinb14e1 (2009-07) 23 Selecting a parameter Ö The name of the parameter e.g.: Unit is displayed as long as the key is pressed or at least fo
24 tinb14e1 (2009-07) 4.5.1 Parameters Switching threshold The PGD400 has a switching function with one adjustable threshold. The status of th
tinb14e1 (2009-07) 25 Pressure unit Unit of measured values, thresholds etc. Value Ö mbar (default) Ö Torr Ö Pascal
26 tinb14e1 (2009-07) 5 Maintenance, Repair The product requires no maintenance. Cleaning the PGD400 For cleaning the outside of the PGD400,
tinb14e1 (2009-07) 27 6 Troubleshooting Signalization of errors Error messages Possible cause and remedy / acknowledge-ment Interruptio
28 tinb14e1 (2009-07) Technical support If the problem persists after the message has been ack-nowledged for several times and/or the gauge ha
tinb14e1 (2009-07) 29 7 Repair Please contact your local INFICON service center. INFICON assumes no liability and the warranty becomes null
tinb14e1 (2009-07) 3 Intended Use The PGD400 is used together with an INFICON Pirani Standard Gauge of the PSG5xx series for total pressure mea
30 tinb14e1 (2009-07) 9 Disposal DANGER DANGER: contaminated parts Contaminated parts can be detrimental to health and environment. Before beg
tinb14e1 (2009-07) 31 Appendix A: Default Parameters Default Usermbar1.0-3 mbar B: Literature [1] www.inficon.com Operating Manual
32 tinb14e1 (2009-07) EC Declaration of Conformity We, INFICON, hereby declare that the equipment mentioned below complies with the provisions o
tinb14e1 (2009-07) 33 Notes
34 tinb14e1 (2009-07) Notes
tinb14e1 (2009-07) 35 Notes
LI–9496 Balzers Liechtenstein Tel +423 / 388 3111 Fax +423 / 388 3700 Original: English tinb14e1 (2009-07) [email protected] t i nb14e1 ww
4 tinb14e1 (2009-07) Contents Product Identification 2 Validity 2 Intended Use 3 Scope of Delivery 3 1 Safety 6 1.1 Symbols Used 6 1.2
tinb14e1 (2009-07) 5 Appendix 31 A: Default Parameters 31 B: Literature 31 EC Declaration of Conformity 32 For cross-references within this
6 tinb14e1 (2009-07) 1 Safety 1.1 Symbols Used DANGER Information on preventing any kind of physical injury. WARNING Information on preventing
tinb14e1 (2009-07) 7 1.2 Personnel Qualifications Skilled personnel All work described in this document may only be carried out by persons who
8 tinb14e1 (2009-07) Disconnecting device The disconnecting device must be readily identifiable and easily reached by the user. To disconnect th
tinb14e1 (2009-07) 9 2 Technical Data Mains specifications Voltage 100 … 240 VAC Frequency 50 … 60 Hz Power consumption ≤30 VA Overvoltage c