tina53e1-c (2011-11) 1 Capacitance Diaphragm Gauge CDG160D CDG200D Operating Manual Incl. EC Declaration of Conformity
10 tina53e1-c (2011-11) 2 Technical Data For further technical data for gauges with DeviceNet and Profibus interface → [5] and [6]. Measur
tina53e1-c (2011-11) 11 Switching functions SP1, SP2 Setting range 0 … 99% F.S. (0 … 9.9 V) Hysteresis 1% F.S. Relay contact 30 VDC/ ≤0.5 A
12 tina53e1-c (2011-11) Supply DANGER The gauge may only be connected to power sup-plies, instruments or control devices that conform to the re
tina53e1-c (2011-11) 13 For longer cables, larger conductor cross-sections are required (Rcable ≤1.0 Ω). Grounding concept → "Power Connec
14 tina53e1-c (2011-11) Dimensions [mm] 8-VCR longfemale54.1 8-VCR female31.531.5DN 16 ISO-KF31.5DN 16 CF-R31.5OD½"1382110.4 3.8 Weight
tina53e1-c (2011-11) 15 Analog Measurement Signal vs. Pressure 12 34567891000.9×F.S.0.5×F.S.0.6×F.S.1.0×F.S.0.7×F.S.0.8×F.S.0.1×F.S.0.2×F.S.0.
16 tina53e1-c (2011-11) 3 Installation WARNING WARNING: fragile components The ceramic sensor may be damaged by impacts. Do not drop the produ
tina53e1-c (2011-11) 17 DANGER DANGER: protective ground Products that are not correctly connected to ground can be extremely hazardous in the
18 tina53e1-c (2011-11) Mount the gauge so that no vibrations occur. The gauge may be mounted in any orientation. To keep conden-sates and par
tina53e1-c (2011-11) 19 Mount the insulation shell. Insulation shell
2 tina53e1-c (2011-11) Product Identification In all communications with INFICON, please specify the infor-mation given on the product nameplat
20 tina53e1-c (2011-11) 3.2 Power Connection Make sure the vacuum connection is properly made (→ 16). DANGER The gauge may only be connec
tina53e1-c (2011-11) 21 3.2.1 D-Sub, 9-pin Connector If no sensor cable is available, make one according to the following diagram (cable lengt
22 tina53e1-c (2011-11) 3.2.2 D-Sub, 15-pin Connector If no sensor cable is available, make one according to the following diagram (cable length
tina53e1-c (2011-11) 23 4 Operation Put the gauge into operation. If you are using an INFICON con-troller, define the measurement range (→ [1
24 tina53e1-c (2011-11) 4.2 Zeroing the Gauge The gauge is factory calibrated while "standing upright" (→ "Calibration Test Repo
tina53e1-c (2011-11) 25 4.2.1 <ZERO> Adjustment The zero can be adjusted via • the <ZERO> button on the gauge, • the diagnost
26 tina53e1-c (2011-11) Operate the gauge for at least 2 hours under constant ambient conditions (until the signal is stable). Briefly pres
tina53e1-c (2011-11) 27 It also permits to adjust an offset of the characteristic curve in order to • compensate for the offset of the measurin
28 tina53e1-c (2011-11) Keep the button depressedmax. ø1.1 mm Push the <ZERO> button again: Fine adjustment within 0...3 s: the zero
tina53e1-c (2011-11) 29 4.3 Switching Functions The two switching functions can be set to any pressure within the measurement range of the gaug
tina53e1-c (2011-11) 3 Validity This document applies to products of the CDG160D and CDG200D series. Part numbers of standard products are ind
30 tina53e1-c (2011-11) 4.3.1 Adjusting the Setpoints The setpoints can be adjusted via • the buttons on the gauge, • the diagnostic port (
tina53e1-c (2011-11) 31 Press the buttonmax. ø1.1 mm For changing the threshold value, push the <ZERO> button and keep it depressed. T
32 tina53e1-c (2011-11) Push the <ZERO> button again: Fine adjustment within 0...3 s: the zero adjustment value changes by one unit Cha
tina53e1-c (2011-11) 33 4.5 Diagnostic Port (RS232C Interface) The diagnostic port <DIAG> permits to output the pressure reading and al
34 tina53e1-c (2011-11) 5 Deinstallation Preconditions: • Vacuum system vented • Vacuum system cooled to <50 °C 5.1 Power Connection
tina53e1-c (2011-11) 35 DANGER DANGER: contaminated parts Contaminated parts can be detrimental to health and environment. Before beginning to
36 tina53e1-c (2011-11) Remove the insulation shell. Insulation shell Remove the gauge from the vacuum system and install the protective li
tina53e1-c (2011-11) 37 6 Maintenance, Repair Under clean operating conditions, the product requires no main-tenance. Gauge failures due to
38 tina53e1-c (2011-11) 8 Disposal DANGER DANGER: contaminated parts Contaminated parts can be detrimental to health and environment. Before b
tina53e1-c (2011-11) 39 9 Accessories Ordering number Communication adapter (2 m) 6) 303-333
4 tina53e1-c (2011-11) If not indicated otherwise in the legends, the illustrations in this document correspond to gauges with D-Sub 15-pin conne
40 tina53e1-c (2011-11) Further Information [1] www.inficon.com Operating Manual Vacuum Gauge Controller VGC032 tinb02e1 INFICON AG, LI–9496
tina53e1-c (2011-11) 41 ETL Certification 3103457 ETL LISTED The products CDG160D and CDG200D comply with the requirements of the following St
42 tina53e1-c (2011-11) EC Declaration of Conformity We, INFICON, hereby declare that the equipment mentioned below complies with the provisions
tina53e1-c (2011-11) 43 Notes
LI–9496 Balzers Liechtenstein Tel +423 / 388 3111 Fax +423 / 388 3700 Original: German tina53d1-c (2011-11) [email protected] t i na53e1-
tina53e1-c (2011-11) 5 Patents EP 1070239 B1, 1040333 B1 US Patents 6528008, 6591687, 7107855, 7140085 Scope of Delivery 1× gauge CDGxxxD
6 tina53e1-c (2011-11) Contents Product Identification 2 Validity 3 Intended Use 4 Functional Principle 4 Trademarks 4 Patents 5 Scope of De
tina53e1-c (2011-11) 7 Further Information 40 ETL Certification 41 EC Declaration of Conformity 42 For cross-references within this document
8 tina53e1-c (2011-11) 1 Safety 1.1 Symbols Used DANGER Information on preventing any kind of physical injury. WARNING Information on preventi
tina53e1-c (2011-11) 9 1.3 General Safety Instructions • Adhere to the applicable regulations and take the necessary precautions for the proce