Pirani Standard Gauge
PSG500/-S, PSG502-S,
PSG510-S, PSG512-S
Operating Manual
Incl. EC Declaration of Conformity
tina44e1-g (2010-06)
Product Identification
In all communications with INFICON, please specify the in-
formation on the product nameplate. For convenient refe-
rence copy that information into the space provided below.
Model:
PN:
SN:
V W
INFICON AG, LI-9496 Balzers
Validity
This document applies to products with the following part
numbers:
PSG500 PSG500-S (
W filament)
350-060 350-080
(DN 16 ISO-KF)
350-062 350-082
(DN 16 CF-R)
350-061 350-081
(1/8" NPT)
350-064 350-084
(8 VCR
®
)
350-065 350-085
(4 VCR
®
)
350-063 350-083
(½"-Rohr)
350-066 350-086
(7/16-20 UNF)
350-067 350-087
(DN 16 ISO-KF long tube)
350-068 350-088
(DN 16 CF-R long tube)
PSG502-S (Ni filament)
350-140
(DN 16 ISO-KF)
350-142
(DN 16 CF-R)
350-141
(1/8" NPT)
350-144
(8 VCR
®
)
350-145
(4 VCR
®
)
350-143
(½"-Rohr)
350-146
(7/16-20 UNF)
350-147
(DN 16 ISO-KF long tube)
350-148
(DN 16 CF-R long tube)
PSG510-S (W filament) PSG512-S (Ni filament)
350-200
(DN 16 ISO-KF) 350-300 (DN 16 ISO-KF)
The part number (PN) can be taken from the product name-
plate.
If not indicated otherwise in the legends, the illustrations in
this document correspond to the gauge with part number
350 060. They apply to gauges with other part numbers by
analogy.
We reserve the right to make technical changes without prior
notice.
All dimensions in mm.
Intended Use
The Pirani Standard Gauges PSG500/-S, PSG502-S,
PSG510-S, PSG512-S have been designed for vacuum
measurement of gases in the pressure range of
5×10
-4
… 1000 mbar.
They must not be used for measuring flammable or combus-
tible gases in mixtures containing oxidants (e.g. atmospheric
oxygen) within the explosion range.
They can be operated in connection with an INFICON con-
troller or with another controller.
Trademark
VCR
®
Swagelok Marketing Co.
Safety
Symbols Used
DANGER
Information on preventing any kind of physical injury.
WARNING
Information on preventing extensive equipment and envi-
ronmental damage.
Caution
Information on correct handling or use. Disregard can lead
to malfunctions or minor equipment damage.
Personnel Qualifications
Skilled personnel
All work described in this document may only be carried out
by persons who have suitable technical training and the
necessary experience or who have been instructed by the
end-user of the product.
General Safety Instructions
Adhere to the applicable regulations and take the nec-
essary precautions for the process media used.
Consider possible reactions between the materials and the
process media.
Consider possible reactions (e.g. explosion) of the process
media due to the heat generated by the product.
Adhere to the applicable regulations and take the neces-
sary precautions for all work you are going to do and con-
sider the safety instructions in this document.
Before beginning to work, find out whether any vacuum
components are contaminated. Adhere to the relevant re-
gulations and take the necessary precautions when hand-
ling contaminated parts.
Communicate the safety instructions to all other users.
Liability and Warranty
INFICON assumes no liability and the warranty becomes null
and void if the end-user or third parties
disregard the information in this document
use the product in a non-conforming manner
make any kind of interventions (modifications, alterations
etc.) on the product
use the product with accessories not listed in the product
documentation.
The end-user assumes the responsibility in conjunction with
the process media used.
Gauge failures due to contamination or wear and tear, as well
as expendable parts (filament), are not covered by the
warranty.
Technical Data
Measurement principle thermal conductance according
to Pirani
Measurement range
(air, O
2
, CO, N
2
)
5×10
-4
… 1000 mbar
Accuracy (N
2
)
1×10
-3
… 100 mbar ±15% of reading
5×10
-4
… 1×10
-3
mbar ±50% of reading
100 … 1000 mbar ±50% of reading
Resolution 1% of reading
Repeatability
1×10
-3
… 100 mbar 2% of reading
Output signal (measure-
ment signal)
Voltage range VDC 0 … +10.3
Measurement range VDC +1.9 … +10.0
Voltage vs. pressure logarithmic
1.286 V/decade
Error signal
Filament rupture
V
V
0 … +0.5
+0.1
Output impedance
Ω
2×4.7
Minimum loaded imped-
ance
kΩ
10, short-circuit proof
Response time ms 80
Gauge identification
27.0 kΩ, referenced to supply
common (voltage at pin 4 5 V)
Adjustment one tactile switch for ATM and
HV adjustment
Switching functions SP1, SP2
Threshold value indi-
cation and setting
one tactile switch at measure-
ment value output. Press briefly
for threshold indication. Keep
pressing or press repeatedly for
threshold setting.
Setting range 2×10
-3
… 500 mbar
Hysteresis 10% above lower threshold
Relay contact
closed
open
30 V, 0.5 ADC, floating
at low pressure (lamp is lit)
at high pressure, error, missing
supply
Supply
DANGER
The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded extra-low vol-
tage (SELV). The connection to the gauge has to
be fused
1)
.
Supply voltage
At gauge VDC +14 … +30
Ripple V
pp
1
Current consumption mA <500
(max. starting current)
Power consumption W 1
Fuse required
1)
AT
(slow)
1
Electrical connection FCC 68 / RJ45 appliance
connector, 8 poles, male
Sensor cable 8 poles plus shielding
Cable length 100 m (8×0.14 mm
2
)
Grounding concept
"Electrical Connection"
Vacuum connection to
signal common
connected via 1 MΩ
(voltage difference <15 V)
Supply common to
signal common
conducted separately, for
differential measurement
Materials exposed to
vacuum
PSG500/-S, PSG502-S
DIN 1.4301, DIN 1.4305,
DIN 1.4435, glass, Ni,
NiFe
PSG510-S,
PSG512-S
Al
2
O
3
(ceramics), Ni,
DIN 1.4435, DIN 1.4305
DIN 1.3981
Filament
PSG500/-S, PSG510-S
PSG502-S, PSG512-S
W
Ni
1)
INFICON controllers fulfill these requirements.
Seitenansicht 0
1 2 3 4

Inhaltsverzeichnis

Seite 1 - Pirani Standard Gauge

Pirani Standard Gauge PSG500/-S, PSG502-S, PSG510-S, PSG512-S Operating Manual Incl. EC Declaration of Conformity tina44e1-g (2010-06) Pr

Seite 2 - Installation

Internal volume DN 16 ISO-KF DN 16 CF-R 1/8" NPT 8 VCR® 4 VCR® ½"-Rohr 7/16-20 UNF cm3 cm3 cm3 cm3 cm3 cm3 cm3 ≈1.5 ≈1.5 ≈2 ≈2 ≈2 ≈2 ≈1.

Seite 3 - Switching Functions

Electrical Connection Make sure the vacuum connection is properly made (→ "Vacuum Connection").  If no sensor cable is available, mak

Seite 4 - 1 copy for file of sender

Spare Parts When ordering spare parts, always indicate: • all information on the product nameplate • description and ordering number according to